6. TECHNICAL SPECIFICATIONS
6.1. TECHNICAL FEATURES
The accuracy is indicated as [% of reading + number of digits]. It is referred to the
following environmental conditions: temperature 23°C ± 5°C, relative humidity < 80%.
DC Voltage
Range Resolution Accuracy Input impedance Overload protection
600.0mV 0.1mV
(0.5rdg+2dgt)
10M //
less of 100pF
1000VDC
750VACrms
6.000V 0.001V
60.00V 0.01V
600.0V 0.1V
1000V 1V
AC TRMS Voltage
Range Resolution
Accuracy (*)
(50 500Hz)
Input impedance
Overload
protection
600.0mV 0.1mV
(0.9rdg+5dgt)
10M //
less then100pF
1000VDC
750VACrms
6.000V 0.001V
60.00V 0.01V
600.0V 0.1V
750V 1V
(*) NOTES
Sinusoidal waveform 50/60Hz only for 600.0mV range
Basic accuracy is specified for sinusoidal waveform up to 4000 counts. Over 4000 counts add 0.6%rdg to the
accuracy
For non-sinusoidal waveform below 2000 counts add 1.5%rdg for crest factor from 1.4 to 3
DC Current
Range Resolution Accuracy Output voltage Overload protection
600.0A 0.1A
(1.0rdg+2dgt)
<4mV/A
600Vrms
6000A 1A
6.000A 1mA
2V max
Fast fuse
10A / 1000V
10A 10mA
AC TRMS Current
Range Resolution
Accuracy
(50 500Hz)
Output voltage Overload protection
6.000A 1mA
(1.5rdg+5dgt)
2V max
Fast fuse 10A /
1000V
10A 10mA
Frequency
Range Resolution Accuracy
Sensitivity (**) Overload protection
6000Hz 1Hz
(0.1rdg+1dgt)
100mVrms (*)
600Vrms
60.00kHz 0.01kHz
600.0kHz 0.1kHz
6.000MHz 0.001MHz 250mVrms
60.00MHz 0.01MHz
1Vrms
(*) for measurements < 20Hz sensitivity is 1.5Vrms
(**) Max sensitivity: <5VACrms