Owner's manual

HT210
EN - 15
6. TECHNICAL SPECIFICATIONS
6.1. TECHNICAL FEATURES
The accuracy is indicated as [% of reading + number of digits]. It is referred to the
following environmental conditions: temperature 18°C(64°F) ÷ 28°C(82°F), relative
humidity < 70%.
DC Voltage (Autorange)
Range Resolution Accuracy Input impedance Overload protection
4.000V 0.001V
±(1.2%rdg+2dgt)
7.8M
600VDC/ACrms
40.00V 0.01V
400.0V 0.1V
600V 1V
±(1.5%rdg+2dgt)
AC Voltage (Autorange)
Range Resolution
Accuracy
(50 ÷ 400Hz)
Input impedance Overload protection
4.000V 0.001V
±(1.2%rdg+4dgt)
7.8M
600VDC/ACrms
40.00V 0.01V
±(1.5%rdg+3dgt)
400.0V 0.1V
600V 1V
±(2.0%rdg+4dgt)
Frequency range: 50Hz ÷ 400Hz
Resistance (Autorange)
Range Resolution Accuracy Overload protection
400.0 0.1 ±(1.2%rdg + 4dgt)
250VDC/ACrms
4.000k 0.001k ±(1.0%rdg + 2dgt)
40.00k 0.01k
±(1.2%rdg + 2dgt)
400.0k 0.1k
4.000M 0.001M
40.00M 0.01M ±(2.0%rdg + 3dgt)
Diode Test
Feature Resolution Accuracy Max open voltage Overload protection
1mV
±(10%rdg + 5dgt)
circa 1.5VDC 250VDC/ACrms
Continuity test
Feature Buzzer Test current Overload protection
<150
<0.3mA 250VDC/ACrms
Frequency (Autorange)
Range Resolution Accuracy Sensitivity Overload protection
5.000Hz 0.001Hz
±(1.5%rdg + 5dgt)
>8Vrms 250VDC/ACrms
50.00Hz 0.01Hz
500.0Hz 0.1Hz
±(1.2%rdg + 3dgt)
5.000kHz 1Hz
50.00kHz 10Hz
500.0kHz 100Hz
5.000MHz 1kHz
±(1.5%rdg + 4dgt)
10.00MHz 10kHz
Duty cycle (Autorange)