Dielectric withstanding test (DIELECTRIC)
Nominal test
voltage UN
(V)
Output
Resolution
(V)
Accuracy of
output voltage
Overvoltage
protection
250 800 COM & 0.250.80 kV
10 ± 3% U
N CAT III 300 V
810 2500 COM & 0.812.50 kV
2510 5100 COM & 2.515.10 kV
Nominal test voltage UN: Adjustable 250 5100 V, 50/60 Hz (floating) in steps of 10 V
Distortion of test voltage: Crest factor = 1.414 ± 5%
Measurement modes: MANUAL, RAMP (timer) or BURN
Output power: 500 VA at 5100 V
Leakage current IAPP:
Display range
(mA)
Resolution
(mA)
Accuracy
0 200
1 ± (3% rdg + 2 mA)
Leakage current IREAL:
Display range
(mA)
Resolution
(mA)
Accuracy
0 110
1 ± (3% rdg + 4 mA)
Nominal breaking current (IAPP or IREAL): Adjustable 1 110 mA in steps of 1 mA
Short-circuit current: > 200 mA
Breaking time: < 30 ms
Note!
Always connect COM terminal to GND if measured OUT is grounded, otherwise possible capacitive
leakage current may flow to ground which may disturb the measurement!