M: Insulation resistance 250, 500VDC (M72, M74, M75)
Range Resolution Accuracy Overload protection
0.00 19.99M 0.01M
(5.0% rdg + 2 dgt)
605V max RMS
20.0 199.9M 0.1M
(5.0% rdg + 2 dgt)
200 999M(*) 1M
(10.0% rdg + 2 dgt)
(*) For 500VDC test voltage. For 250VDC test voltage the range is: 200 499M
Autorange
Open circuit voltage: <1.3 x V
0
Accuracy of nominal voltage: –0% +10%
Short circuit current: <3.0mA
Nominal testing current: 1mA @ 1K x V (1mA @ 500K)
RCD: test on RCD devices AC type (M73, M74, M75)
Range Resolution Accuracy Overload protection
2 400ms
1ms
(2.0% rdg + 2 dgt)
605V max RMS
Phase to ground voltage: 110 265V
Test currents: 30mA, 30mA x 5, 100mA, 300mA
Frequency: 50Hz 0.5Hz / 60Hz 0.5Hz
Ra : Measurement of global earth resistance (M73, M74, M75)
Test current Range Resolution Accuracy Overload protection
15mA
1 1999 1
(5% rdg + 2 dgt)
605V max RMS
100mA
0.1 199.9 0.1
(5% rdg + 3 dgt)
Phase to ground voltage: 110 265V; Frequency: 50Hz 0.5Hz / 60Hz 0.5Hz
Nominal voltage used for the calculation of the prospective short circuit current: 127V if 100V V
measured
< 150V
230V if 150V V
measured
< 265V
Wire mapping (M75)
Length of the cable: 1100m
Number of remote units: max 8 units
Detected errors: OPEN Pairs, REVERSED pairs, SHORT pairs, SPLIT pairs, CROSSED pairs, MISWIRING
According to the norm: TIA568B
CAUTION
In all measurements the instrument dispalys the symbol when:
The instrument is operating in a critical situation as for example in
presence of overvoltages.
The instrument cannot guarantee the measuring uncertainty lower than
30% of the reading, according to IEC/EN61557-1.