User manual

Gx Series
EN - 96
10.2. REFERENCE GUIDELINES
Safety: IEC/EN61010-1, IEC/EN61557-1, -2, -3, -4, -5, -6, -7, -10
Technical documentation: IEC/EN61187
Safety of measuring accessories: IEC/EN61010-031, IEC/EN61010-2-032
Insulation: double insulation
Pollution level: 2
Max operating altitude: 2000m (6562ft)
Protection index: IP40
Measurement category: CAT III 240V to earth, maximum 415V between inputs
LOW (200mA): IEC/EN61557-4
M: IEC/EN61557-2
RCD: IEC/EN61557-6 (only on Phase-Neutral-Earth systems)
LOOP P-P, P-N, P-PE: IEC/EN61557-3
EARTH: IEC/EN61557-5
123: IEC/EN61557-7
Multifunction: IEC/EN61557-10
Short circuit current : EN60909-0
Earth resistance TN system: EN61936-1 + EN50522 (not for USA, Germany, Extra
Europe countries)
10.3. GENERAL CHARACTERISTICS
Mechanical characteristics
Size (L x W x H): 225 x 165 x 75mm ; (9 x 6 x 3in)
Weight (batteries included): 1.2kg ; (42 ounces)
Power supply
Battery type: 6x1.5V alkaline batteries type AA IEC LR06 MN1500
6 x1.2V rechargeable batteries NiMH type AA
Low battery indication: low battery symbol “ ” on the display
Battery life: > 500 tests for each function
Auto Power OFF: after 5 minutes idling (if activated)
Miscellaneous
Display: TFT, color, capacitive touch-screen, 320x240mm
Memory: 999 memory locations, 3 levels of markers
Connection to PC: optical/USB port
Remote connection: WiFi connection
10.4. ENVIRONMENT
10.4.1. Environmental conditions for use
Reference temperature: 23° ± 5°C ; (73°F ± 41°F)
Operating temperature: 0 ÷ 40°C ; (32°F ÷ 104°F)
Allowable relative humidity: <80%RH
Storage temperature: -10 ÷ 60°C ; (14°F ÷ 140°F)
Storage humidity: <80%RH
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC
(LVD) and of EMC Directive 2004/108/EC
10.5. ACCESSORIES
See the attached packing list.