User manual

Gx Series
EN - 61
CAUTION
Possible additional earth connections could influence the measured value. In
case of real difficulty in removing them, we recommend performing the
measurement in an indirect way.
5. The value of the measured leakage current appears in
real time on the display as shown in the screen to the
side.
Press the SAVE button or touch the
icon to save
the measurement (see ยง 7.1).