2. GENERAL SPECIFICATIONS
DISPLAY AND MEMORY:
Features: Touch screen, color graphic LCD, 320x240mm
Memory: 999 locations, 3 marker levels
Communication: Optical-USB and built-in WiFi
POWER SUPPLY:
Batteries: 6 x 1.2V(rechargeable) type AA or 6 x 1.5V type AA
Battery life: > 500 test for each funtions
Auto Power OFF: after 5 min of idleness (disabled)
MECHANICAL FEATURES:
Dimensions (L x W x H): 225 x 165 x 75mm
Weight (included batteries): 1.2kg
WORKING ENVIRONMENTAL CONDITIONS:
Reference temperature: 23°C 5°C
Working temperature: 40°C
Allowed relative humidity: < 80% HR
Storage temperature: -10 60°C
Storage humidity: < 80% HR
TEST VERIFIES REFERENCE STANDARDS:
Continuity test with 200mA: IEC/EN61557-4
Insulation resistance: IEC/EN61557-2
Fault loop impedance: IEC/EN61557-3
RCD test: IEC/EN61557-6
Phase sequence: IEC/EN61557-7
Multifunction: IEC/EN61557-10
Prospective short circuit current: EN60909-0
Earth resistance on TN systems: EN61936-1 + EN50522
GENERAL REFERENCE STANDARDS:
Safety of measuring instruments: IEC/EN61010-1, IEC/EN61010-031, IEC/EN61010-2-032
Product type standard: IEC/EN61557-1
Technical documentation : IEC/EN61187
Insulation: double insulation
Pollution degree: 2
Encapsulation : IP40
Overvoltage category: CAT III 240V (to ground), max 415V between inputs
Max height of use: 2000m
This instrument complies with the requirements of the European Low Voltage Directives 2006/95/EEC
(LVD) and EMC 2004/108/EEC