Datasheet

COMBIG2 Rel 1.00 of 27/08/14
Professional installation safety tester
Pag 1 of 6
HT ITALIA SRL
Via della Boaria 40 - 48018 Faenza (RA)- Italy
Tel: +39-0546-621002 - Fax: +39-0546-621144
email: export@htitalia.it - web: http://www.ht-instruments.com
1. ELECTRICAL SPECIFICATIONS
Accuracy is indicated as ( readings + no. of digits*resolution) at 23°C 5°C, <80HR
Voltage (RCD, LOOP, Phase sequence)
Range [V] Resolution [V] Accuracy
15 460
1
(3.0 rdg + 2dgt)
Frequency
Range [Hz] Resolution [Hz] Accuracy
47.0 63.6
0.1
(0.1 rdg + 1dgt)
Continuity test on protective and equalizing conductors
Range [] Resolution []
Accuracy (*)
0.01 19.99
0.01
(5.0rdg + 3dgt)
20.0 99.9
0.1
(*) calibrate the cables to null their resistance
Test current: > 200mA DC for R5 (calibration included) ; Resolution for DC current :1mA
Open-circuit voltage: 4V V
0
12V
Insulation resistance (DC voltage)
Test voltage[V]
Range [M] Resolution [M]
Accuracy
50
0.01 9.99
0.01
(2.0rdg + 2dgt)
10.0 49.9
0.1
50.0 99.9
0.1
(5.0rdg + 2dgt)
100
0.01 9.99
0.01
(2.0rdg + 2dgt)
10.0 99.9
0.1
100.0 199.9
0.1
(5.0rdg + 2dgt)
250
0.01 9.99
0.01
(2.0rdg + 2dgt)
10.0 99.9
0.1
100 499
1
(5.0rdg + 2dgt)
500
0.01 9.99
0.01
(2.0rdg + 2dgt)
10.0 199.9
0.1
200 499
1
500 999
1
(5.0rdg + 2dgt)
1000
0.01 9.99
0.01
(2.0rdg + 2dgt)
10.0 199.9
0.1
200 999
1
1000 1999
1
(5.0rdg + 2dgt)
Open-circuit voltage: nominal test voltage –0% +10%
Short circuit current: <6.0mA at 500V test voltage
Nominal test current: >1mA if load= 1k*Vnom (Vnom=50V, 100V, 250V, 500V, 1000V)
Safety protection: the display shows an error message for input voltage >10V
Z Line (Line-Line, Line-Neutral, Line-PE)
Range [] Resolution []
Accuracy
0.00 199.9 m (*)
0.1
m (*)
(5.0 rdg + 1
m) (*)
200 1999 m (*)
1
m (*)
0.01 9.99
0.01
(5.0 rdg + 3dgt)
10.0 199.9
0.1
(*) By means of IMP57 optional accessory
Maximum test current: 5.81A (at 265V); 10.10A (at 457V)
Test voltage ranges: 100265V (Line-Neutral) / 100460V (Line-Line); 50/60Hz 5%
Protection type: MCB (B, C, D, K), Fuse (gG, aM)
Insulation materials: PVC, Rubber butyl, EPR, XLPE
First fault current (IT systems)
Range (mA) Resolution (mA) Accuracy
0.1 0.9
0.1
(5.0 rdg + 1dgt)
1 999
1
(5.0 rdg + 3dgt)
Limit contact voltage (ULIM) : 25V, 50V

Summary of content (6 pages)