HP MSA 1040 User Guide (762783-001, March 2014)
60 Troubleshooting
When a power supply sensor goes out of range, the Fault/ID LED illuminates amber and an event is
logged to the event log.
Power supply module voltage sensors
Power supply voltage sensors ensure that the enclosure power supply voltage is within normal ranges.
There are three voltage sensors per power supply.
Table 23 Controller module temperature sensor descriptions
Description Normal operating
range
Warning operating
range
Critical operating
range
Shutdown values
CPU temperature 3°C–88°C0°C–3°C,
88°C–90°C
> 90°C0°C
100°C
FPGA temperature 3°C–97°C0°C–3°C,
97°C–100°C
None 0°C
105°C
Onboard temperature 1 0°C–70°CNoneNoneNone
Onboard temperature 2 0°C–70°CNoneNoneNone
Onboard temperature 3
(Capacitor temperature)
0°C–70°CNoneNoneNone
CM temperature 5°C–50°C ≤ 5°C,
≥ 50°C
≤ 0°C,
≥ 55°C
None
Table 24 Power supply temperature sensor descriptions
Description Normal operating range
Power Supply 1 temperature –10°C–80°C
Power Supply 2 temperature –10°C–80°C
Table 25 Voltage sensor descriptions
Sensor Event/Fault LED condition
Power supply 1 voltage, 12V < 11.0 0 V
> 13.00V
Power supply 1 voltage, 5V < 4.00V
> 6.00V
Power supply 1 voltage, 3.3V < 3.00V
> 3.80V