- Hewlett-Packard Portable Generator - Generator User Manual
TB 9-6625-2049-35 
18 
 (6) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.742 and 
3.032 V ac. 
 (7) Press keys and enter values using DATA keys as listed in (a) through (c) below: 
 (a) FUNCTION positive ramp. 
 (b) 100 Hz. 
 (c) AMPTD CAL. 
 (8) Multimeter will indicate between 2.843 and 2.931 V ac. 
 (9) Enter 499 Hz using DATA keys. Multimeter will indicate between 2.843 and 
2.931 V ac. 
 (10) Enter 501 Hz using DATA keys. Multimeter will indicate between 2.598 and 
3.176 V ac. 
 (11) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.598 and 
3.176V ac. 
 b. Adjustments. No adjustments can be made. 
17. Dc Offset 
 a. Performance Check 
 (1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50Ω feedthrough 
termination. 
 (2) Press SIGNAL key off (option 002 not lit). 
 (3) Press presently active FUNCTION key to remove ac output and activate 
ENTRY DC OFFSET key (lit). 
 (4) Enter 5 VOLT using DATA keys and press AMPTD CAL key. 
 (5) Multimeter will indicate between 4.98 and 5.02 V dc. 
 (6) 
Enter 
-5 VOLT
 using 
DATA 
keys. Multimeter will indicate between -4.98 
and -5.02 V dc.
 (7) Repeat technique of (6) above using TI settings and multimeter indications 
listed in table 5. 
Table 5. Dc Offset 
Test instrument  
DC OFFSET 
Multimeter indications  
(V dc) 
settings  Min  Max 
 1.499 VOLT  1.493  1.505 
 -1.499 VOLT  -1.505  -1.493 
 499.9 mV  0.4979  0.5019 
 -499.9 mV  -0.5019  -0.4979 
 149.9 mV  0.1493  0.1505 
 -149.9 mV  -0.1505  -0.1493 
 49.99 mV  0.04979  0.05019 
 -49.99 mV  -0.05019  -0.04979 