Troubleshooting Tables
4-17
9030-9035, 9042-
9045, 9050, 9051,
9069-9085
Direct transmission protocol error. Retry transmission.
9038 A non-specific error occurred during transmission. Retry transmission at 2400 bps.
9046 Remote device not supported. Verify that the remote site is compatible with the
PageWriter cardiograph.
9047-9049 Requested an ECG from an unrecognized device. Verify that the remote site is a TraceMaster ECG
Management System.
9052-9065 Premature quit command received from remote site. Contact remote site.
9066 Defective FLASH memory 1. Perform the extended self-test (refer to Chapter
2).
2. If the extended self-test fails, replace the Option
#A05 PCA.
9067 Memory allocation fault during transmission. 1. Turn off, wait 30 seconds, turn on.
2. Retry transmission.
9086 Unrecognized device for transmission Verify that the remote site is compatible with the
PageWriter cardiograph.
9087-9089 Printer problem during transmission. 1. Check the paper supply.
2. Be sure the printer door is closed.
3. Retry the transmission.
9163 Transmission stopped unexpectedly.
x of n ECGs sent
Remote site stopped communication
Press any key to continue
or
Transmission stopped unexpectedly.
x of n ECGs received
Remote site stopped communication
Press any key to continue
1. Verify the transmission speed matches that of the
remote site.
2. Verify the remote system communication is oper-
ating correctly
9200 An ECG was requested from a FAX machine. Select a connection type of Modem, Direct, Modem-
SCP, or DirectSCP to request an ECG from a remote
site.
9500-9674 Unable to build an SCP record for transmission. Turn off, wait 30 seconds, turn on. Retry the trans-
mission. If the error persists, call HP.
9575-9895 SCP transmission errors. Retry the transmission. Call HP if the error persists.
9585, 9586 The ECG transmission stalled before completion,
either because the file transfer protocol is absent or
is in an unexpected state, or another communication
error occurred.
Wait one minute, then retry the transmission. Check
or restart the communication software at the remote
site. If the error persists, call HP.
9640, 9769 The receiving device rejected a transmitted ECG
because of data errors.
Wait one minute, then retry the transmission. If the
error persists, call HP.
Event # Definition Corrective Action