Service manual
Troubleshooting Tables
4-15
In addition to the error codes listed above, the following event codes are displayed on the
extended self-test report for cardiographs equipped with Option #A05.
Table 4-9 Event Codes (Option #A05 only)
Event # Definition Corrective Action
1223 PageWriter XLi record corrupted during transmis-
sion. This code is followed by an sub-error code.
Retry the transmission.
3573 1. The battery was removed when the cardiograph
was not plugged in to AC power.
2. The battery fuse is defective.
Replace the battery.
3575 The cardiograph turned off due to inactivity. Normal event. No action required.
3577 The cardiograph turned off due to low battery. Plug cardiograph into AC power.
3700 The power was cycled quickly. Normal event. No action required.
3701 Errors were detected in the front end communication
link.
Run the extended self-test to check operation of the
front-end.
5160, sub code 3 No modem at remote site. 1. Check all cable connections.
2. Be sure the remote modem is turned on and con-
nected to the telephone cable.
3. Check the cardiograph telephone directory and
verify telephone number, device type, and trans-
mission speed.
4. Check the modem initialization string commands
that allow compatibility with the remote modem.
5. Be sure the remote site is ready to receive the
transmission. Consider extending the time the
modem waits for a connection by adding the
string “S7=90” to the modem initialization string.
6. Retry the transmission
5160, sub code 4 Check modem configuration. Check the modem initialization string for legal
modem commands. See your modem documentation
for legal commands for your modem.
5160, sub code 5 Check telephone cable. Check the telephone cable connection to the
modem. Attach the telephone cable to a telephone to
verify a dial tone on the line.
5160, sub code 6 Telephone busy, waiting to redial...
(after 30 seconds, this message appears:)
Telephone busy, redialing...
Normal event. The cardiograph will redial, waiting
30 seconds between attempts, until the line is open
or until the operator stops the transmission.
5160, sub code 7 No answer, waiting to redial...
(after 30 seconds, this message appears:)
No answer, redialing...
Be sure the remote site is ready to receive a trans-
mission. If necessary, extend the time the modem
will wait for a connection by adding the string
“S7=90” to the modem initialization string.