Service manual

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix
2-27
System Safety (S)
Use Safety Analyzer
Check:
Chassis-to-Ground Resistance
Groundwire Leakage Current
(Normal condition)
Groundwire Leakage Current
(Single Fault)
Patient Lead Leakage Current
•Source (Normal Condition)
•Source (Single Fault Condition)
•With Mains on applied part
(Single Fault Condition)
•Auxiliary (Normal Condition)
•Auxiliary (Single Fault Condi-
tion)
Indicate Applicable Safety
Standard as follows:
Note: All leakage current test are
Normal and Reverse Polarity
Conditions
For the above - record the worst
case value for multiple lead
measurements.
<
200 m aaa
<
500 uA (<300uA UL) bbb
<
1000 uA cccc
<
10 uA dd
<
50 uA ee
<
50 uA (< 20 uA UL) ff
<
10 uA gg
<
50 uA hh
i=I for IEC i
i=U for UL
S:aaa,bbb,ccc,dd,
ee,ff,gg,hh,i
example:
S:09,400,900,8,4
0,30,8,40,I
Note: When recording test results, separate tests by a semi-colon(;). For example: V:p;PO:p;F:p;
D:p;L:p;P:p;K:p;ECG:p;S:90,400,900,8,40,30,8,40,I
Table 2-10 Performance Verification: PageWriter 300pi/200i/200/100 (M1770A, M1771A, M1772A)
Test Block Name
Test or "Inspection" to
Perform
Expected (Passing)
Test Results
Data to
Record
x=p (pass) or
f (fail)