Service manual

Resetting the Cardiograph to the Factory Default State
2-17
2Resetting the Cardiograph to the Factory Default State
Resetting the cardiograph sets the Auto ECG report counter to 0, and resets the
PageWriter 200/200i/300pi configuration to the factory default.
Resetting the PageWriter 200/200i/300pi
To reset the PageWriter 200/200i/300pi perform the following steps:
1. Using Table 2-9, turn the cardiograph On in a language different from the current lan-
guage.
2. The cardiograph is now reset to the factory default state.
3. If it is necessary to operate the cardiograph in the original language, turn the cardio-
graph to Standby, then turn the cardiograph On using the appropriate key combina-
tion from Table 2-9.
Resetting the PageWriter 100
To reset the PageWriter 100 perform the following steps:
1. Turn the cardiograph to Standby.
2. Press and hold and while turning the cardiograph on.
3. Turn the cardiograph to Standby.
4. Press and hold and while turning the cardiograph on.
5. The cardiograph is now reset to the factory default state.