Service manual

Specification Data
1-5
1
Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Meets
frequency response standard 3.2.7.2 using methods A, D, and E with auto report filter settings at 0.05-150 Hz or
0.15-150 Hz.
2
-4 dB @ 100 Hz
3
Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Meets
frequency response standard 3.2.7.2 using methods A, D, and E with manual report filter settings at 0.05-150 Hz.
4
Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices.
Table 1-3 ECG to Paper Specifications
Parameter Specification
Frequency response of PageWriter
200/200i/300pi
(-3dB bandwidth)
High Pass Low Pass
Auto ECG
0.05 Hz
1
0.15 Hz
1
0.5 Hz
150 Hz
1
100 Hz
2
40 Hz
Manual ECG
0.05 Hz
3
0.15 Hz
0.5 Hz
150 Hz
3
100 Hz
2
40 Hz
Frequency response of PageWriter 100
Auto ECG
1
Manual ECG
3
<3 dB down at (user selectable):
<
3 dB down at (user selectable):
0.15-150 Hz or 0.15-40 Hz
0.05-150 Hz or 0.05-40 Hz
ECG resolution on paper
voltage axis
time axis
Approximately 8 dots/mm (200 dots/in)
Approximately 20 dots/mm (500 dots/in) at 25mm/sec and slower
Approximately 10 dots/mm (250 dots/in) at 50mm/sec
Recorder speed 5, 10, 25, and 50 mm/sec +
1.5%
4
ECG visibility with pace pulse Meets or exceed AMMI EC11-1991 standard for Diagnostic Electrocardio-
graphic Devices