User Manual

Usage guidelines
Please note the following tips and guidelines when using the X-Rite i1Display Pro with the display:
The display may take up to ve seconds after instrument connection to display the rst calibration
menu. This is normal for this instrument.
This instrument should be used in direct connection mode with the counterweight adjusted to maintain
the instrument position during calibration.
HP recommends that calibration be conducted in a dimly lit environment.
The ambient light lter must be rotated away from the instrument lens prior to measurements being
made. If it is left in place, measurement errors will occur.
The i1Display Pro has an error tolerance for factory calibration of each instrument. Visible calibration
dierences can occur if a dierent probe is used for each display. HP strongly recommends that a single
i1Display Pro kit be used to calibrate all displays within a facility or workow.
Colorimetry Research CR-250 spectroradiometer
The Colorimetry Research CR-250 is a low-cost spectroradiometer than can be used as either a contact
instrument with the optional rubber hood or as a distance measurement device using an optional viewing
system.
Conguration
No conguration is required for the CR-250 instrument.
Usage guidelines
Please note the following tips and guidelines when using the Colorimetry Research CR-250 with the display:
The display may take up to 15 seconds after instrument connection to display the rst calibration menu.
This is normal for this instrument.
If used as a contact instrument, HP recommends calibration in a dim lighting environment.
If used as a noncontact instrument, HP strongly recommends calibration in a dark environment.
70 Chapter 3 Display calibration