User Manual
Usage guidelines
Please note the following tips and guidelines when using the X-Rite i1Display Pro with the display:
● The display may take up to ve seconds after instrument connection to display the rst calibration
menu. This is normal for this instrument.
● This instrument should be used in direct connection mode with the counterweight adjusted to maintain
the instrument position during calibration.
● HP recommends that calibration be conducted in a dimly lit environment.
● The ambient light lter must be rotated away from the instrument lens prior to measurements being
made. If it is left in place, measurement errors will occur.
● The i1Display Pro has an error tolerance for factory calibration of each instrument. Visible calibration
dierences can occur if a dierent probe is used for each display. HP strongly recommends that a single
i1Display Pro kit be used to calibrate all displays within a facility or workow.
Colorimetry Research CR-250 spectroradiometer
The Colorimetry Research CR-250 is a low-cost spectroradiometer than can be used as either a contact
instrument with the optional rubber hood or as a distance measurement device using an optional viewing
system.
Conguration
No conguration is required for the CR-250 instrument.
Usage guidelines
Please note the following tips and guidelines when using the Colorimetry Research CR-250 with the display:
● The display may take up to 15 seconds after instrument connection to display the rst calibration menu.
This is normal for this instrument.
● If used as a contact instrument, HP recommends calibration in a dim lighting environment.
● If used as a noncontact instrument, HP strongly recommends calibration in a dark environment.
70 Chapter 3 Display calibration