5.7.3 Calibration signal
The torque transducer delivers an electrical shunt signal that can be activated
by using switch (6) on the evaluation unit. See also 5.7.5
The transducer should not be under load when the shunt signal is being
measured, as the shunt signal is mixed additively.
5.7.4 Stator antenna
The cable connection between stator antenna and evaluation unit is realized
with a coaxial cable.
NOTE
• When you connect the stator antenna, please take regard to a stress
relief
• Please tighten the connection screws with the appropriate tightening
torque.
• Use thread locker (medium strength, e.g. LOCTITE No. 242) to glue the
screws into the counter thread to exclude pre-stressing loss due to screw
slackening.
• Allign the stator like shown in the picture below