6.3 Supply voltage
The transducer must be operated with a separated extra‐low voltage (supply
voltage 9…36 VDC).
7 Functional testing
You can check the functionality of the rotor and the stator from the LEDs on
the evaluation unit.
It is not allowed to extend or reduce the length of the connection
cable between rotor and evaluation unit!!!
Don’t lay cable parallel to high voltage and control cable. If this is not
avoidable, please let a minimum gap of 50cm and lay the cable inside a
steel tube.
Avoid transformer, motor, gate, thyristor circuits and similarly leakage
fields.
Important
Check the calibration signal specified on the type label or test protocol.
To obtain stable conditions, the measurement should be started or the
calibration signal should be activated only once the transducer has been
warming up for 15 minutes.