6 General information
With extension cables, make sure that there is a proper connection
with minimum contact resistance and good insulation.
All cable connectors or swivel nuts must be fully tightened
.
6.2 EMC protection
The product offered is a special assembly for stationary systems that is not
available on the general market or a transducer for installation by system
integrators and plant manufacturers. According to EMVG
1
§12 paragraph 2 and
Directive 2004/1008/EC article 13 paragraph 1 this product does not require an
EC declaration of conformity nor the CE marking.
This product is intended exclusively for subsequent processing by companies or
persons that are experts in the field of electromagnetic compatibility (EMC).
Relevant EMC protection requirements relating to the product offered are met
when the following Installation notes are observed and implemented.
Installation notes
Please note the following points during installation and use:
It is essential to observe the specifications and notes provided in the
operating manual and the data sheet
Connecting cables, in particular the measuring and control cables, need to be
shielded
Make sure that the transducer and shielding are connected extensively to
ground.
Ensure an interference-free environment, avoid radiation interference
Devices connected to this product need to comply with protection
requirements per EMVG (Gesetz über die elektromagnetische Verträglichkeit
von Geräten / law on electromagnetic compatibility of instruments).
You have to install the shield of the connection cable at the shielded housing of the
electronics, to achieve the EMC-protection of the measuring chain. Make sure that the
transducer and shielding are connected extensively to ground.
Important