User's Manual

Table Of Contents
Load‐carrying capacity
T40B A3452-10.0 53
9 Load‐carrying capacity
Nominal (rated) torque can be exceeded statically up to
the limit torque. If the nominal (rated) torque is exceeded,
additional irregular loading is not permissible. This
includes longitudinal forces, lateral forces and bending
moments. Limit values can be found in Chapter 13
“Specifications”, page 58.
Measuring dynamic torque
The torque flange can be used to measure static and
dynamic torque. The following apply to the measurement
of dynamic torque:
S The T40B calibration performed for static
measurements is also valid for dynamic torque
measurements.
S The natural frequency f
0
of the mechanical measuring
arrangement depends on the moments of inertia J
1
and J
2
of the connected rotating masses and the
torsional stiffness of the T40B.
Use the equation below to approximately determine the
natural frequency f
0
of the mechanical measuring
arrangement:
f
0
+
1
2p
· c
T
·
ǒ
1
J
1
)
1
J
2
Ǔ
Ǹ
f
0
= natural frequency in Hz
J
1,
J
2
= mass moment of inertia in kgVm
2
c
T
= torsional stiffness in NVm/rad
S The permissible mechanical oscillation width
(peak‐to‐peak) can also be found in the specifications.