User's Manual
Table Of Contents
- T40B
- T40B
- English
- 1 Safety instructions
- 2 Markings used
- 3 Application
- 4 Structure and mode of operation
- 5 Mechanical installation
- 6 Electrical connection
- 7 Shunt signal
- 8 Functionality testing
- 9 Load‐carrying capacity
- 10 Maintenance
- 11 Waste disposal and environmental protection
- 12 Ordering numbers, accessories
- 13 Specifications
- 14 Supplementary technical information
- English
- T40B
- Deutsch
- 1 Sicherheitshinweise
- 2 Verwendete Kennzeichnungen
- 3 Anwendung
- 4 Aufbau und Wirkungsweise
- 5 Mechanischer Einbau
- 6 Elektrischer Anschluss
- 7 Shuntsignal
- 8 Funktionsprüfung
- 9 Belastbarkeit
- 10 Wartung
- 11 Entsorgung und Umweltschutz
- 12 Bestellnummern, Zubehör
- 13 Technische Daten
- 14 Ergänzende technische Informationen
- 15 Dimensions | Abmessungen
- 15.1 T40B without rotational speed measurement | T40B ohne Drehzahlmessung
- 15.1.1 T40B 50 Nm - 100 Nm, Option 7, Code S
- 15.1.2 T40B 50 Nm, 100 Nm, Option 7, Code U
- 15.1.3 T40B 200 Nm, Option 7, Code S
- 15.1.4 T40B 200 Nm, Option 7, Code U
- 15.1.5 T40B 500 Nm - 1 kNm, Option 7, Code S
- 15.1.6 T40B 500 Nm - 1 kNm, Option 7, Code U
- 15.1.7 T40B 2 kNm - 3 kNm, Option 7, Code S
- 15.1.8 T40B 2 kNm - 3 kNm, Option 7, Code U
- 15.1.9 T40B 5 kNm, Option 7, Code S
- 15.1.10 T40B 5 kNm, Option 7, Code U
- 15.1.11 T40B 10 kNm, Option 7, Code S
- 15.1.12 T40B 10 kNm, Option 7, Code U
- 15.2 T40B with rot. speed measurement and reference signal | mit Drehzahlmessung und Referenzimpuls
- 15.2.1 T40B 50 Nm - 100 Nm, Option 7, Code S
- 15.2.2 T40B 50 kNm - 100 Nm, Option 7, Code U
- 15.2.3 T40B 200 Nm, Option 7, Code S
- 15.2.4 T40B 200 Nm, Option 7, Code U
- 15.2.5 T40B 500 Nm - 1 kNm, Option 7, Code S
- 15.2.6 T40B 500 Nm - 1 kNm, Option 7, Code U
- 15.2.7 T40B 2 kNm - 3 kNm, Option 7, Code S
- 15.2.8 T40B 2 kNm - 3 kNm, Option 7, Code U
- 15.2.9 T40B 5 kNm, Option 7, Code S
- 15.2.10 T40B 5 kNm, Option 7, Code U
- 15.2.11 T40B 10 kNm, Option 7, Code S
- 15.2.12 T40B 10 kNm, Option 7, Code U
- 15.1 T40B without rotational speed measurement | T40B ohne Drehzahlmessung
- Deutsch
Functionality testing
T40B A3452-10.0 51
8.1 Rotor status, LED A (upper LED)
Color Significance
Green (pulsating) Internal rotor voltage values ok
Flashing orange
Rotor and stator mismatched (an increasing flashing
frequency indicates the degree of misalignment)
=> Correct the rotor/stator alignment
Pulsating orange
Rotor status cannot be defined
=> Correct the rotor/stator alignment
If the LED still pulsates orange, it is possible that there is a
hardware defect. The measurement signals reflect the level
of the fault.
Red (pulsating)
Rotor voltage values not ok.
=> Correct the rotor/stator alignment
If the LED still pulsates red, it is possible that there is a
hardware defect. The measurement signals reflect the level
of the fault.
Pulsating means that the LED goes dark for about 20 ms
every second (sign of life), making it possible to detect
that the transducer is functioning.