User's Manual
Table Of Contents
- T40FH
- T40FH
- English
- 1 Safety instructions
- 2 Markings used
- 3 Application
- 4 Structure and mode of operation
- 5 Mechanical installation
- 6 Electrical connection
- 7 TEDS transducer identification (Option 3, Code PNJ)
- 8 Shunt signal
- 9 Functionality testing
- 10 Load-carrying capacity
- 11 Maintenance
- 12 Waste disposal and environmental protection
- 13 Dimensions
- 14 Ordering numbers, accessories
- 15 Specifications
- 16 Supplementary technical information
- English
- T40FH
- Deutsch
- 1 Sicherheitshinweise
- 2 Verwendete Kennzeichnungen
- 3 Anwendung
- 4 Aufbau und Wirkungsweise
- 5 Mechanischer Einbau
- 6 Elektrischer Anschluss
- 7 Aufnehmer‐Identifikation TEDS (Option 3, Code PNJ)
- 8 Shuntsignal
- 9 Funktionsprüfung
- 10 Belastbarkeit
- 11 Wartung
- 12 Entsorgung und Umweltschutz
- 13 Abmessungen
- 14 Bestellnummern, Zubehör
- 15 Technische Daten
- 16 Ergänzende technische Informationen
- Deutsch
Electrical connection
44 A4429-1.0 HBM: public T40FH
7 TEDS transducer identification
(Option 3, Code PNJ)
TEDS stands for "Transducer Electronic Data Sheet". An
electronic data sheet can be stored in the transducer as
defined in the IEEE1451.4 standard, making it possible
for the amplifier to be set up automatically. A suitably
equipped amplifier reads out the transducer characteris
tics (electronic data sheet), translates them into its own
settings and measurement can then start.
The digital identification system is available at plug con
nection PIN 7 to PIN 2. The HBM TEDS Editor is used to
store the data. This is a component of the HBM
"MGCplus Setup Assistant" software. You can use the
Editor to manage different user rights, thus protecting the
essential transducer data from being overwritten by mis
take.
7.1 Hierarchy of user rights
7.1.1 Standard rights (USR level)
This level concerns rights which the user of the trans
ducer needs in order to change the entries which depend
on the conditions of use.
7.1.2 Calibration rights (CAL level)
This level concerns rights which are needed by a calibra
tion laboratory, for instance, if the sensitivity in the TEDS
memory needs to be changed.