User's Manual

32
13
Specifications
Type
T40HS
Accuracy class
0.05
Torque measuring system
Nominal (rated) torque M
nom
N
m
500 1000 2000 3000
Nominal (rated) sensitivity
(range between
torque = zero and nominal (rated) torque)
Frequency output 10/ 60 / 240 kHz kHz 5
1)
/ 30
2)
/ 120
3)
Voltage output V 10
Sensitivity tolerance (deviation of the actual
output quantity at M
nom
from the nominal
sensitivity)
%
± 0.1
Output signal at torque = zero
Frequency output kHz 10
1)
/ 60
2
)
/ 240
3
)
Voltage output V 0
Output signal at torque = zero
Frequency output
with positive nominal (rated) torque
with negativ nominal (rated) torque
Voltage output
with positive nominal (rated) torque
kHz
kHz
V
15
1)
/ 90
2)
/ 360
3)
( 5V symmetrical
4)
)
5
1)
/ 30
2)
/ 120
3)
( 5V symmetrical
4)
)
+10…+12
with negative nominal (rated) torque V -10…-12
Load resistance
Frequency output
k
2
Voltage output
k
10
Long
-
term drift over 48 h
Frequency output %
<
±
0.03
Voltage output %
<
±
0.03
Measurement frequency range
(
-
3dB )
1
1)
/ 3
2
)
/ 6
3
)
Group delay
µs < 400
1)
/ < 220
2
)
/ < 150
3
)
Residual ripple
Voltage output
5
)
mV < 40
Effect of t
emperature per 10 K in the
nominal (rated) temperature range
on the output signal,
related to the actual
value of signal span
Frequency output %
<
±
0.05
Voltage output %
<
±
0.2
on the
zero signal, related to the nominal
(rated) sensitivity
Frequency output %
<
±
0.05
Voltage output %
<
±
0.1
1) Option 5, 10 ± 5kHz ( Code SU2 )
2) Option 5, 60 ± 30kHz ( Code DU2 )
3) Option 5, 240 ±120kHz ( Code HU2 )
4) RS-422 complementary signals, observe terminating resistance
5) Signal frequency range 0.1 to 10 kHz