User's Manual

10
4
Mechanical installation
4.1 Important
precautions
during
installation
NOTE
A T40HS torque flange is a precision measurement element and therefore needs careful
handling. Dropping or knocking the transducer may cause permanent damage. Make
sure that the transducer cannot be overloaded, even while it is being mounted.
Handle the transducer with care.
Check the effect of bending moments, critical speeds and natural torsional
oscillations, to prevent the transducer being overloaded by increases in
resonance.
Make sure that the transducer cannot be overloaded.
W
ARNING
There is a danger of the transducer breaking if it is overloaded. This can cause danger for
the
operating personnel
of the system in which the
transducer
is
installed
.
Implement appropriate safety measures to avoid
overloads and to protect against resulting
dangers
.
Use a screw locking device (medium strength, e.g. LOCTITE No. 242) to glue the
screws into the counter thread to exclude prestressing loss due to screw
slackening, if alternating loads are to be expected.
Comply with the mounting dimensions to enable
correct operation.
An appropriate shaft flange enables the T40HS torque flange to be mounted directly. It is
also possible to mount a joint shaft or relevant compensating element directly on the rotor
(using an intermediate flange when required). Under no circumstances should the
permissible limits specified for bending moments, lateral and longitudinal forces be
exceeded. Due to the T40HS torque flange's high torsional stiffness, dynamic shaft train
changes are kept to a minimum.