User's Manual

26
Assignment for connector
2:
Speed output signal (optional).
Device
connector
2
5
4
3
8 1
7 6
Top view
Plug
Pin
Assignment
1
Measurement signal speed
1)
(pulse string, 5 V; 0°)
2
Not in use
3
Measurement signal speed
1)
(pulse string, 5 V; phase shifted by 90°)
4
Not in use
5
Not in use
6
Measurement signal speed
1)
(pulse string, 5 V; 0°)
7
Measurement signal speed
1)
(pulse string, 5 V; phase shifted by 90°)
8
Supply voltage zero
Shielding connected to housing ground
1) RS-422 complementary signals; with cable lengths exceeding 10 m, we recommend
using a
termination resistor of R = 120 ohms.
Assignment for connector
3:
Supply voltage and frequency output signal.
Device
connector
6 1
5 7 2
4
3
Top view
Plug
Pin
Assignment
1
Torque measurement signal (volt. output; 0 V
)
2
Power supply 0 V;
3
Supply voltage 18 V - 30 V DC
4
Torque measurement signal (voltage output,
±10
V)
5
Not in use
6
Shunt signal resolution 5 V - 30 V
7
Shunt signal 0 V;
Shielding connected to housing ground
Assignment for connector
4:
TMC
-
only for connection to the TIM 40 Torque Interface Module within HBM.
Version b 13.01.2015 HBM: public