Technical Manual
XNX Universal Transmitter
Section 2 - Installation and Operation
72
days (or fewer, in accordance with customer site procedures) to
assure the highest level of safety.
Use the
switches to highlight the current interval and use
✓
to select it.
Figure 107. Edit Interval and Setting Interval Value Screens
Use the switches to move to the desired position. Use
✓
to
select it. Use the
switches to decrement or increment the value
untilthedesiredvalueisreached.(Theminimumnumberofdaysis0;
themaximumnumberisdenedbythesensortype.)Use
✓
to select
thevalueandmovetothenexteld.Repeatforeacheld.Whenall
oftheeldshavebeenupdated,usethe switches to highlight
the
‘ü’
on the right side of the display. Use
✓
on the front panel to
save the settings.
Caution: Setting the Calibration Interval to zero turns off the
calibration notification. This can seriously affect sensor performance.
Figure 108. Saving New Interval and New Interval Accepted Screens
✓
Accept New Sensor Type
When replacing EC cells or mV sensors, use Accept New Sensor
Type to load default parameters into the XNX transmitter for
calibration and sensor life. Accept New Sensor Type is also used
when replacing an EC cell with another EC cell for a different
target gas. (See Section 4.2.2).
Figure 109. Accept New Sensor Type Menu
When changing the target gas by inserting a new sensor, the
transmitterwillprompttheuserforaconrmationofthechange
before adjusting to the properties of the new sensor.
Figure 110. Select New Sensor
In the display of the XNX transmitter, the old sensor type and
the new sensor type will be displayed. Use
✓
to accept the new
sensor or
✖
to reject it.
Beam Block Options
The Beam Block Options menu is available only if the XNX
transmitter is connected to a Searchline Excel sensor.