Install Instructions

TP970-72, TP970-74, TP9600, AND TP9630 PNEUMATIC SENSORS, HUMIDISTATS, AND THERMOSTATS
Piping
Figures 46 and 47 show typical one-pipe and two-pipe
applications.
S TAT
M
VALVE
C3159
Fig. 46. Typical One-Pipe Application.
S TAT
VALVE
M
BM
C3160
Fig. 47. Typical Two-Pipe Application.
CALIBRATION
Tools Required
Calibration requires the following tools, shown in Figure 48:
Thermostat Tool CCT735A (MQT735A)
Gage 305965 (0 to 30 psi [0 to 210 kPa])
Gage Port Needle CCT729 (MQP729)
Digital Relative Humidity Indicator Pen CCT915 for
humidistats and humidity sensors
Test Thermometer CCT902 (MQT902) for thermostats
and temperature sensors
C3929
GAGE PORT NEEDLE
DIGITAL RELATIVE
HUMIDITY INDICATOR PEN
100
120
140
160
180
80
60
200
220
40
20
0
10
25
30
20
15
5
0
GAGE
TEST THERMOMETER
THERMOSTAT TOOL
Fig. 48. Humidistat/Humidity Sensor Calibration Tools.
Humidistat/Humidity Sensor
General
After installing the humidistat, set the humidistat to the
desired setpoint and let the system operate long enough to
stabilize. The length of time required for stabilization depends
on system response time and could be only a few minutes or
as long as several hours. Verify that the system has stabilized
before checking calibration.
NOTE: Humidistats and humidity sensors are factory
calibrated and should require only a bleed-off check
to ensure correct operation. The throttling range is
factory set and should not require adjustment under
normal operating conditions.
45
95-5597—1