Product Manual

Table Of Contents
Honeywell BW Solo User Manual
31
9.16. Passcode
By default, no passcode is required for access to the instrument’s settings. However, a 4-digit
passcode can be set, which will prevent access to all settings. This can be set here or through
Honeywell SafetySuite software. If a passcode is ever lost and the instrument cannot be accessed,
contact Technical Support.
Click once to advance the number (0, 1, 2, 3, etc.).
Press and hold the button for 1 second to advance to the next digit.
When the last digit is reached, press and hold the button for 1 second.
Confirm by clicking the button once.
Important! If you forget your passcode, then contact Honeywell Technical Support.
9.17. Always On
Enabling the Always On feature is prevents the user from turning the instrument off during their
shift. When Always On is active, this icon is displayed, along with the words “Always On”:
There are two ways to disable it:
1. Enter the Settings menu, go to Always On, and toggle it off.
2. Insert the instrument into an IntelliDox and disable it through Safety Suite Device Configurator.
To turn the BW Solo off, you can disable Always On or insert the instrument into an IntelliDox. After
there is no IR communication, it will turn the unit off in 6 seconds
If you have turned it off through the IntelliDox, then the next time it is turned on the Always On
feature will still be enabled.
IMPORTANT!
Always On reduces battery life to 1/3 of expected full life because instead of running for 8 hours per
day, it runs 24 hours per day.
9.18. Exit
Exit to the main menu