Commissioning
SmartRadar FlexLine Part No.: 4417.762_Rev07
6 - 12 Service Manual
6.3.3.4 Commisioning
For a correct functioning of the HCI-GPU module within an instrument
(gauge), the following entities can be set by using either Engauge or
SmartView.
By using the following table, check each entity for correctness.
Name Value Range Default Value Explanation
[FlexConn Modbus protocol switch] <FlexConn-GPU>
<Modbus>
<FlexConn-GPU> If production-, test-,
configuration-, and/or
service tools are to be
used, switch to the
FlexConn protocol by
selecting
<FlexConn-GPU>.
[RS Baudrate] <1200>
<2400>
<4800>
<9600>
<19200>
<38400>
<57600>
<115200>
<19200>
Communication speed
[Turn around delay] <0..2000 ms> <0 ms> The turnaround delay is the
minimum time the HCI-
GPU waits, before starting
answering the host.
[Parity] <Odd>
<Even>
<None>
<Odd>
User can have different
parity depending upon
application.
<Stopbits> <One>
<Two>
<One>
[Modbus slave address] <1..247> <1> On a site there can be
multiple gauges connected
via the RS-485 physical link
and hence different CAN-
RS boards can be identified
by unique slave addresses.
However, note that RS-485
supports only 32 devices
in multidrop and so must be
the addresses.