User Manual

Service Tools
Part No.: 4417.762_Rev07 SmartRadar FlexLine
Service Manual 4 - 1
CHAPTER 4 SERVICE TOOLS
4.1 SmartView
4.1.1 General
The SmartView is the basic tool with which the user can communicate
with the SmartRadar FlexLine modules.
As normally open keyboard contacts may be dangerous in an
explosion-hazardous environment, the SmartView is built up as a totally
shielded explosion-safe tool.
4.1.2 SmartView Versions
The SmartView can be delivered in 3 versions:
SmartView fixed on the instrument
SmartView as a tank-side indicator (stand-alone)
A portable SmartView
ESF07-0027
FIGURE 4-1 The 3 different SmartView options
4.1.3 Connections
Within a hazardous environment, connecting or disconnecting electrical
equipment is dangerous, because of sparking risks.
The portable SmartView however is designed such that it may be
connected/disconnected within a hazardous zone.
The fixed version SmartViews (integrated within the instrument or tank-
side mounted) have pre-installed fixed connections.
fixed on the instrument as a tank-side indicator as a portable tool