Part No.: 4417.762_Rev07 SmartRadar FlexLine
Service Manual 3 - 7
Finally the status code informs about the specific reason why the status
is good, uncertain, or bad. This information is presented as an
information number coupled with a textual description of this specific
situation. See table below.
Each board implements “health” as generic information based on the
function (s) health(s).
Each function implements “health” as generic information.
Each sensor and digital I/O function implements an entity called
“Primary Value”, which in addition to the actual measured value also
contains a “health” status.
Status Status Category
Good
actual
manual
last valid
stored
Uncertain
instrument
environment
Bad
general hardware fail
general firmware fail
general commissioning fail
general calibration fail
general operational fail
over range (data available!)
under range (data available!)
no data available
un-initialized
killed
001 056 077 104 632 328 207 782 065 478 199 389 011 072
........................................................
manual stored
instrument
environment
last
valid
actual
un-init
killed
no data
under range
over range
hw fail
sw fail
cal fail
Good Uncertain Bad