User Manual

Commissioning
Part No.: 4417.762_Rev07 SmartRadar FlexLine
Service Manual 6 - 147
6.10.3.2.2 Vapour Temperature
NOTE: Using Engauge, following entities are set within the
Engauge Vapour temperature tab.
Some Vapour temperature settings are shared with
Product temperature settings.
The following entities can be set by Engauge or SmartView for a correct
functioning of the FII-VT module in an instrument.
Check each entity for its correctness.
Name Explanation Value Range Default
[Gas immersion depth] The minimum required
distance below an element
before it is taken into
account in the average
vapour temperature
calculation
floating point number:
<-x.x .. +x.x>
<0.5>
[High High alarm]
[High Alarm]
[Low Alarm]
[Low Low alarm]
4 thresholds for activating a
related alarm status in the
Primary Value
floating point numbers:
<-x.x .. +x.x>
<+1.0E22>
[Alarm test enable] Enables (if activated, see
next listed entity) the
simulation of one of the 4
alarms for a minute, by
simulating the actual
measured Primary Value is
below or above the alarm
threshold.
<enable>
<disable>
<disable>
[Alarm test]
Activates and selects at the
same time the Alarm test (if
enabled with the [Alarm test
enable]
listed before, and the
4 alarm thresholds are
properly set).
[No Alarm]
[High High alarm]
[High Alarm]
[Low Alarm]
[Low Low alarm]
<No Alarm>
[Function identification]
The current module’s
function name. This
function is visible on the
SmartView display.
13 characters <Vapour temp.>