User Manual
Commissioning
SmartRadar FlexLine Part No.: 4417.762_Rev07
6 - 130 Service Manual
*) Default values for entities marked in GREEN in the
above table could be considered as proper values for
making the health for respective Function or Board
<GOOD>. However, the user is free to choose these
values as per the requirement along with the rest of the
entities.
All the entities in the above table would be initialized to their default
values after
[init novram] command is given. All the entities should have
their proper values so as to make the respective Function and Board
health to
<GOOD> status.
The
[Analog Output at 4 mA] and [Analog Output at 18 mA] fields should be
entered after giving proper calibration commands (for more details, see
6.9.4).
[HART PV unit code] Unit code for linked PV <UNIT_CODE_UNKNOWN> Function 1 Communication
(HART
communication)
[HART Upper Transducer
Limit]
Upper transducer limit for
connected sensor of linked
PV board
<0>
[HART Lower
Transducer Limit]
Lower transducer limit for
connected sensor of linked
PV board
<0>
[Polling Address] Polling address of the
device, which is used for
poll-based addressing
<0>
[Transducer Serial
Number]
Transducer serial number
of connected sensor of
linked PV board
<123>
[Upper Range Value] Upper range value for PV.
This value is used to
calculate the analog output
current.
<0> Function 2 Sensor
(Analog Output)
[Analog Output at 4mA] Calibration reading when
[Calibrate at 4 mA] command
is given
<0>
[Analog Output at 18
mA]
Calibration reading when
[Calibrate at 18 mA] command
is given
<0>
[Lower Range Value] Lower range value for PV.
This value is used to
calculate the analog output
current.
<0>
[Analog Burnout Value] Analog burnout value in
case of malfunctioning
device
<3.6mA>
Name Explanation Default Value
Function
Part
Function