User Manual
Commissioning
SmartRadar FlexLine Part No.: 4417.762_Rev07
6 - 116 Service Manual
TABLE 6-12 PV unit type and function identification
6.8.3.7 SmartView Display
Although the Primary Values (PV) of functions 1 through 8 are
calculated in SI units, the SmartView display supports the following
units for local calibration, commissioning etc. The displayed unit
depends on the value of the
[PV selected unit] entity.
Function Allocation PV Unit Type
Function
Identification
Function 1 P1 Pressure UNIT_TYPE_PRESSURE
<P1 Pressure>
Function 2 P3 Pressure UNIT_TYPE_PRESSURE <P3 Pressure>
Function 3 Distance UNIT_TYPE_LENGTH <Distance>
Function 4 Temperature UNIT_TYPE_TEMPERATURE <Temperature>
Function 5 Pressure UNIT_TYPE_PRESSURE <Pressure>
Function 6 Density UNIT_TYPE_DENSITY <Density>
Function 7 Other UNIT_TYPE_UNDEFINED <Other Units>
Function 8 HIMS Density UNIT_TYPE_DENSITY <HIMS Density>
PV Unit
Type
Display Unit
Type
Displayed
Units
PV
Selected
Unit
Range
Pressure Pascal
<pas> 1 max. 0.99 MPa
Kilo Pascal
<kpas> 2 max. 9.9 MPa
Pounds per square
inch (small)
<psi s> 3 max. 99 psi
Pounds per square
inch (large)
<psi l> 4 max. 999 psi
Temperature Degrees Celsius
<°C> 1-
Degrees Fahrenheit
<°F> 2-
Distance Meters
<m> 1-
Feet
<ft> 2-
Inches
<in> 3-
Fractions
(feet, inches,
1/16
th
inch)
<f.i.s.> 4-
Density Kilograms per cubic
meter
<kg/m3> 1-
Pounds per cubic
foot
<lb/f3> 2-
Degrees API
<a.p.i> 3-