User Manual

8
3.The security valve and the exhaust steam valve should be sent to qualified organization to check regularly to ensure the safety.
4.The volume of the inner cylinder can not be more than 4/5 of the total volume of the instrument.
5. Make sure the sealing ring not contact the oil to avoid steam loss because of the broken adhesive tape.
6. Things which are easy to explode when contacts the steam or boost abruptly during the sterilization are both strictly prohibited.
7.When you want to sterilize the liquid,it is advisable that use a durus Heat resistant glass wares to put the liquid,and the volume of
the liquid can not be more than 3/4 of the wares,it is prohibited to use a wrapped rubber or cork at the mouth of the wares,it is advisable to
use a cotton plugs.
8.It is prohibited to put sterilized goods(wound-dressing and liquids)with different type and different sterilizing requirement
together into one instrument.
9.If the pressure indicator shows not right after long time use,you should have it serviced regularly
10.The security valve should also be examined regularly to ensure the reliability,if the security valve not take off when the
working pressure above 0.17MPa,then this security valve should be examined and replaced,it is easy to arise accident
because the security valve can not release the pressure well and make the inside of cylinder too much of the pressure.
11.The fuses should also be examined regularly,you should also replace it after long time use,if the instrument doesn’t
supply well when you connected to the power,then you should check if the plug and the socket or the fuses is broken
or not.
12.transport and storing environment
A.required temperature: 4055
B.required relative humidity: 80%
C.required atmospheric pressure: 5001060hPa
13. operating environment
A.required temperature: +540
B.required relative humidity: 75%
C.required atmospheric pressure: 5001060hPa
Maintenance of the product
1. Examine the sealing effectiveness of all pipes and connections regularly.
2. Please check if the instrument be adequately grounded before you start to use the machine again after stop using of the machine more than one
month.
3. Examine regularly to ensure the reliability of the sealed ring and flannel not drop out locally, please put the sealed ring into the water trough
promptly to avoid sudden pressure releasing accidents because of sealed ring dropped out during the process of increasing the pressure.
PACKING LIST
1.ONE INSTRUCTION
2. ONE Warranty card
3.ONE CERETIFICATE
.Failure handling
ITEM
Symptom of instrument failure
analysis of causes
solution
1
1
The time of filling into the water level regulator
significantly extended
2
The water indicator still show water shortage
1
The water level regulator is
blocked up with some foreign
bodies
3
Clear the pipes or the
detector.
4
Replace promptly.