Operation Manual

74
17. Interfering Substances depend on the concentration. The below
substances up to the test concentration will not affect the test
results:
Regularity and Safety Notice
WEEE note
The WEEE (Waste Electrical and Electronic Equipment) Directive,
that came into effect as European law on 13th February 2003, led
to a major change in the disposal of electrical equipment.
The primarily purpose of this directive is to prevent electronic was-
te (WEEE), while encouraging for reuse, recycling and other forms
of reprocessing to reduce waste.
80 mg/dL
(4.4 mmol/L)
250 mg/dL
(13.9 mmol/L)
500 mg/dL
(27.8 mmol/L)
Ascorbic Acid 4 mg/dL (0.26 mmol/L) 10.89% -1.76% 4.55%
Ibuprofen 50 mg/dL (2.43 mmol/L) 3.10% 2.88% 4.62%
L-Dopa 1.8 mg/dL (0.09 mmol/L) 10.59% 7.91% 4.90%
Sodium Salicylate 50 mg/dL (3.12 mmol/L) -2.59% 9.42% -0.84%
Tetracycline 1.5 mg/dL (0.03 mmol/L) -5.32% 3.81% 3.20%
Tolbutamide 100 mg/dL (3.70 mmol/L) -2.60% 12.30% 0.89%
Bilirubin-
unconjugated
2.4 mg/dL (0.04 mmol/L) -2.52% 4.05% -0.23%
Uric acid 8 mg/dL (0.48 mmol/L) 2.71% 9.55% -1.75%
Xylose 4 mg/dL (0.27 mmol/L) -5.12% -1.64% -4.44%
Concentrations
of the interferance tested
Bias
Glucose
Level