Optional: Control Solution Testing
Running a control test lets you know that your meter and test strips
are working properly to give reliable results. You should run a con-
trol test when:
• You use the smartLAB
®
global W nG blood glucose meter for the
first time.
• You open a new vial of test strips.
• You think the meter or test strips might be working incorrectly.
• You drop the meter.
• You have repeated a test and the test results are still lower or
higher than expected.
• You are practicing the test procedure.
Note: Professional users are instructed to follow federal, state, and
local guidelines concerning QC practices.
About the smartLAB
®
nG control solution
• smartLAB
®
nG control solution is not intended for human
consumption or injection.
• Use only with smartLAB
®
nG test strips.
• Write the date you first opened the bottle on the bottle label.
• The smartLAB
®
nG
control solution is durable for three months
from the date the vial is opened or until the „Expiration Date“
on the bottle, whichever comes first.
• Do not use smartLAB
®
nG
control solution that is past the
„Expiration Date“
• The smartLAB
®
nG
control solution can stain clothing. If you
spill it, wash your clothes with soap and water.