Operation Manual

52
Function check
You can carry out a function check with the Check Strip which is
included in your meter set. You do not have to do the function
check before every measurement. It helps you to control if your
smartLAB
®
nG meter works properly, from time to time, though.
1. Insert check strip into the meter:
Make sure your check strip is inserted with writing to top. The meter
will turn on automatically and the display will show „CHK“.
2. Check message in LCD display:
The function check will take approx. 3 seconds. The check can pro-
duce the following messages on the display:
• „OK” - your meter works properly
• „FAL“- change batteries and check again. If the message still ap-
pears, please refer to your local supplier.
After showing “OK” the device will start to send measured data. If
device has not been set yet by the system it will be set now. After-
wards the data will be shown on the screen.
3. Remove check strip:
The meter will turn off automatically after removing the check strip.
Please keep the check strip in the soft case of the meter.