Operation Manual

51
ENGLISH
III. Setup & Operating procedures
Installing batteries
The meter uses 2 CR 2032 3V Lithium
batteries. Batteries will normally last
for aprox. 500 tests. Other types of CR
3V lithium batteries are also acceptable,
yet the capacity of test times may differ.
Install the batteries, by removing the plastic protection when you
first use the meter or replace with new batteries when the “LP” mes-
sage and the low battery symbol appear on the LCD display. Open
the battery lid on the back side of the smartLAB
®
global W nG meter
and insert batteries with plus sign to top. First insert the battery on
the right side and then press the left side down until it clicks. Close
the battery lid. The display will light up and all symbold will blink
for around 10 seconds. Afterwards you can use the device.
Note:
1. Remove the batteries when not using the meter for one month
or longer.
2. Values won’t be deleted when the batteries are replaced.
3. The device needs to be set again after battery change. Please
make sure that the device is close to a corresponding receiving de-
vice for settings.
Meter, batteries, lancets, test strips etc. must be disposed of
according to local regulations at the end of their usage.