Table 2 Guidance and Manufacturer declaration - electromagnetic immunity -
for all ME Equipment and ME System
Guidanceandmanufacturer’sdeclaration–electromagneticimmunity
Thedeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.
The customer or the user of the device should assure that it is used in such an
environment
IMMUNITY test IEC 60601 test level Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
Ifoorsarecoveredwith
synthetic material, the
relative humidity should be
at least 30%.