Datasheet

7
Easy integration into automatic testing equipment (RM3544-01)
High-speed, comprehensive productivity
support
The RM3544-01 delivers the speed demanded by automatic test-
ing equipment at a sophisticated level. The entire process from the
start of measurement to outputting of the judgment result takes as
little as 18 ms. One cycle of operation, lasting from measurement
to judgment output, completes within this time.
The RM3544-01 supports RS-232C data communications at up to
115.2 kbps.
The instrument’s USB interface can also be used.
The EXT I/O output mode can be switched between judgment
mode and BCD mode.
Functionality for verifying the EXT I/O con-
nection status and testing EXT I/O
In addition to allowing you to check EXT I/O signal input on the in-
strument’s screen, this functionality allows you to turn output signals
on or off as desired. This capability simplifi es veri cation work dur-
ing PLC programming.
Handler (EXT I/O) interface
The handler interface (EXT I/O) is isolated from measurement cir-
cuitry, control circuitry, and the protective ground (chassis ground),
providing a high level of noise resistance.
Measurement time
*1
Measurement speed (ms)
FAST
MED SLOW
50Hz 60Hz
21 18 101 401
Tolerance: ±10% ±2 ms
*1 With TC set to ON and the comparator set to ON
t0 : Trigger pulse ON time (0.1 ms or more)
t1 : Trigger pulse OFF time (1 ms or more)
t2 : Measurement start time (max. 1 ms)
t3 :
Capture processing time ; FAST(50Hz): 20.0 ms, FAST(60Hz): 16.7 ms, MEDIUM: 100 ms, SLOW: 400 ms
t4 : Calculation time ; 1 ms
Example of Typical EXT I/O Timing (EOM output hold)
ON
t0
t1
ON
ON
OFF
OFF
OFF OFF
OFF OFF
OFF OFFON/OFF
t3t2
t4
Measurement
processing
TRIG
INDEX
EOM
HI,IN,LO
ERR,BCDm-n
A switch on the rear panel is used to toggle the input signal polarity between NPN
(sink output support) and PNP (source output support) settings depending on the
PLC common polarity.
EXT I/O Input and Output Circuits
EXT I/O Electrical Specifi cations
Inputs:
Photocoupler isolation: Non-voltage contact inputs
(support for current sink output)
Input ON:
Residual voltage: Max. 1 V @4 mA
Input OFF:
Open
Max. 100
μA
Outputs:
Photocoupler-isolated open drain output (no-polarity)
DC30Vmax, DC50mAmax/ch
Residual voltage: Max. 1 V @50 mA, or 0.5 V @10 mA
Internal isolated power supply:
Output voltage
:
Sink output support: 5.0V±10%,
Source output support: -5.0V±10%
Max. output current
:
100mA
Example of Typical
Input Signals
TRIG : External trigger
0ADJ : Zero-Adjust
PRINT : Print
KEY_ LOCK : Key-Lock
BCD_LOW
: Lower digit speci cation when
set to BCD output
LOAD0 to LOAD3
: Panel number to load
IN0, IN1 : General-purpose input pins
Output Signals
HI, IN, LO : Comparator Hi, IN, LO
EOM : End of Measurement
INDEX : End of Import
ERR : Measurement Fault Output
HILO
: Outputs HI or LO when set to
BCD output.
BCDm-n
: Outputs the nth bit of the mth
digit when set to BCD output.
OUT0 to OUT2
: General-purpose output pins
when set to judgment mode
RNG_OUT0 to
RNG_OUT3
: Outputs range information when
set to BCD output
.
ISO_5V : Internally Isolated 5 V
ISO_COM : Internally Isolated Common
When designing a control system using the
EXT I/O interface, be sure to read
the instruction manual and check the neces-
sary technical information.
EXT I/O test function screen
Output Circuit
RM3544-01
ISO_COM
Internally Isolated
Common
Max. 50 mA DC
Zener voltage
= 30 V
10Ω
Output
Input Circuit(Case of NPN type)
RM3544-01
Input
ISO_COM
Internally Isolated
Common
1kΩ
NPN
EXT I/O
MODE
SELECTOR
PNP
2kΩ
During processing