User Manual
41
────────────────────────────────────────────────────
3.5 Settings on the Waveform Display Screen (REC&MEM)
────────────────────────────────────────────────────
NOTE
3.5 Settings on the Waveform Display Screen (REC&MEM)
Time values output to the printer and displayed on the screen with the
additional recording function enabled are equal to those of the most recently
acquired waveforms. Therefore, when measuring waveforms in a different
time axis range, always take that difference into consideration.
When the 10000 (128 M words: 40000, 512 M words: 160000) DIV has
been reached, old data will be overwritten.
In the recorder, the trigger mark ( ) is written as the start position mark. In
an additional recording, the trigger mark is displayed in front of the most
recently entered data.
If auto-save is enabled, only newly acquired waveforms are saved. Even
when A and B cursors appear on the screen, no partial save is performed,
because the setting is disabled when the
START
key is pressed.
A one-line data appears between the two measurements to delimit them.
(The line of data is vertical.)
When using the 8953-10 HIGH RESOLUTION UNIT, 8954
VOLTAGE/TEMP UNIT, or 8955 F/V UNIT, and its additional recording
function is active, if logic input is activated during measurement, a logic
waveform is displayed, but the data is invalid. When measurement is made
with additional recording ON, the screen shows the logic waveform obtained
before additional recording was turned on, but the data is invalid. If logic
input is active, the analog data corresponding to that logic channel becomes
12-bit data. For details,
see Quick Start Section 6.3.10.
Explains the setting items on the Waveform display screen.
For details on setting,
refer to Section 3.3
.
When want to use the Jog/Shuttle control, press the
VALUE
select key. (The
selection window is not displayed.)
Pressing the F4 function key "REC&MEM" on the Waveform display screen
toggles screen between memory waveform and recorder waveform.
REC&mem
Recorder waveform display in Recorder and Memory function
rec&MEM
Memory waveform display in Recorder and Memory function
REC&MEM
Displays both the recorder waveform and memory waveform