User Manual
40
────────────────────────────────────────────────────
3.4 Setting the Additional Recording Function
────────────────────────────────────────────────────
Additional recording: OFF
Initial recording
Records 30 divisions of data
Additional recording: ON
Initial recording
Records 30 divisions of data
10000 DIV
(
No memor
y
ex
p
ansion
)
30 DIV
10000 DIV
(
No memor
y
ex
p
ansion
)
30 DIV
Second recording
Records a further 30 divisions of data
Initial measurements are erased, and new
data is recorded from the beginning of
the memory.
Second recording
New data is recorded from the end of
existing data without erasing the previou
s
measurements.
30 DIV
30 DIV
The first and second sets of
waveforms can be observed by
scrolling or printing the waveform.
30 DIV
Function
display
Meaning
:
Additional recording is disabled.
:
Additional recording is enabled.
3.4 Setting the Additional Recording Function
This records, regarding the memory as though it were recording paper.
As waveforms captured by the recorder are stored in the instrument's
memory, this machine can be operated as a paper recorder.
The memory capacity of 32 M words can accommodate up to 10000
divisions of waveform data. With 128 M words (after optional memory
expansion), waveform data of up to 40000 divisions can be saved. With 512
M words (after optional memory expansion), waveform data of up to 160000
divisions can be saved.
The waveform can be scrolled and printed out.
Switching the additional recording on and off affects the use of memory as
shown below.
(When the recording length is set to 30 divisions)
Procedure Screen: STATUS
1. Press the
STATUS
key to display the STATUS screen.
2. Move the flashing cursor to the
Record Add
item.
3. Use the function keys to make a setting.