2.7 9322 DIFFERENTIAL PROBE Connection
────────────────────────────────────────────────────
DANGER
When using grabber clips, the 9322's maximum rated voltage to
earth is 1500 V AC or DC (CAT ll) / 600 V AC or DC (CAT lll);
when using alligator clips, it is 1000 V AC or DC (CAT ll) / 600 V
AC or DC (CAT lll). To avoid electrical shock and possible
damage to the instrument, never apply voltage greater than
these limits between the input channel terminals and chassis,
or across the input of two 9322s.
Maximum input voltage is 1000 V AC/2000 V DC (CAT ll) / 600 V
AC or DC (CAT lll). Attempting to measure voltage in excess of
the maximum rating could destroy the instrument and result in
personal injury or death.
H
L
Grabber clips
H-L
2,000 V DC,
1,000 V AC (CAT II)
600 V AC, DC (CAT III)
H-GND
1,500 V AC, DC (CAT II)
600 V AC, DC (CAT III)
H
L
Alligator clips
H-L
2,000 V DC,
1,000 V AC (CAT II)
600 V AC, DC (CAT III)
H-GND
1,000 V AC, DC (CAT II)
600 V AC, DC (CAT III)
L-GND
1,500 V AC, DC (CAT II)
600 V AC, DC (CAT III)
L-GND
1,000 V AC, DC (CAT II)
600 V AC, DC (CAT III)
2.7 9322 DIFFERENTIAL PROBE Connection
9322 is a differential probe that connects to input of 8855 MEMORY
HiCORDER input module. For more details, refer to its instruction manual.