User Manual

19
────────────────────────────────────────────────────
2.4 Connection of the Input Product
────────────────────────────────────────────────────
External covering
Internal covering Thermocouple wire
10 mm
25 mm
Recommended diameter
φ
0.4 mm to 1.2 mm
(AWG26 to 16)
NOTE
How to remove the covering from the thermocouple
To install and remove the thermocouple, use a tool such as a screwdriver.
The push-button terminal block on the 8954 VOLTAGE/TEMP UNIT is for
use with a thermocouple. Only connect the thermocouples specified (types
K, J, E, T, N, R, S, B, and W).
If you reverse the + and connections on the thermocouple, the correct value
will not be displayed.