User Manual
APPENDIX 12
────────────────────────────────────────────────────
Appendix 3 Reference
────────────────────────────────────────────────────
X-Y waveform
Shaded area is calculated
No enclosed range,
Line area is calculated
(14) X-Y area value
Calculates the area (V
2
) after X-Y plotting.
The waveform is plotted on the X-Y screen, and the area enclosed by the
plot lines is calculated.
In single, dual, or quad screen, the A/B cursors (vertical, trace) can be used
to specify the range (see Quick Start Section 8.2.) for X-Y plotting and area
calculation.
On the X-Y screen of the memory function, it is not possible to specify the
range with the A/B cursors.
(15) Specified level time
Searches from the top of the arithmetic range to the point intersecting the
specified level to find the time from that point to the trigger.
(16) Pulse width
Calculates the time difference between the point where the rising or falling
waveform intersects the set level and the intersection of the next opposite
slope.