User Manual
189
────────────────────────────────────────────────────
10.6 Printing Procedure
────────────────────────────────────────────────────
10.6.4 Partial Print
NOTE
This function prints the waveform between the A and B cursors (vertical or
trace cursors). The function is available also when the A/B cursors are
currently outside the range displayed on screen. Printing is possible also
when the print format is currently set to "numeric". For details regarding the
use of the A/B cursors, refer Quick Start to Section 8.2.
Memory function
Specified range (out of entire data recorded from a measurement) is printed.
Recorder function
Specified range (out of last 20,000 (expended 128 M is 50,000, 512 M is
200,000) divisions (magnification: x 1) of data in memory) is printed.
Procedure
Screen: Waveform display
1. Press the
DISP
key to display the Waveform
display screen.
2. Position cursor A at the start point of the range
to be printed.
3. Specify the end point of the range with cursor B.
(Move cursor B (or A) to the right. When the
cursor is at the rightmost edge of the screen, the
waveform scrolls to the left, and cursor A (or B)
scrolls with it.)
4. Then press the
PRINT
key. The specified range
is printed, also if cursor A (or B) is currently off
screen. When only the cursor A is used, the
waveform data from the position of cursor A to
the end of the data is printed.
On the waveform evaluation screen, cursors A and B are available, but
partial printing cannot be performed.
On the XY screen, the partial printing cannot be performed.