User Manual
186
────────────────────────────────────────────────────
10.6 Printing Procedure
────────────────────────────────────────────────────
10.6.1 Manual Print
NOTE
10.6 Printing Procedure
This mode serves to print waveform data from the internal memory.
Memory function
Measurement data from one measurement (entire recording length) are
printed.
Recorder function
Measurement data stored in memory before the end of measurement are
printed (2000 divisions
*1
max.).
*1
: Expanded to 128 M words : 50000 DIV, 512 M words : 200000 DIV,
Procedure
Screen: Waveform display
After the measurement is completed, press the
PRINT
key.
Since data are stored, they can be printed as often as desired.
When magnification/compression was used, the printout reflects this
condition.
If you want to stop the printing, press the
STOP
key.
When A-B cursor is ON, it prints that area.When executing from the Print
screen (system), you can select the all screens Between A-B cursor if the
Print Selection when Executing is Yes.
On the recorder, when the recording length is set to Continuous and the time
value display of the Print screen (system) is set to Time, the time value
display is not the time from the triggering point (or the start of recording)
when recording over 20000 DIV (128 Mwords is 50000 DIV, 512 Mwords
is 200000 DIV).When set to Date, it prints the time when printing starts.