User Manual

168
────────────────────────────────────────────────────
9.1 Waveform GO/NG Evaluation (MEM, FFT Function, Power Monitor Function)
────────────────────────────────────────────────────
NOTE
Trigger mode: SINGLE
Measurement continues until stop mode conditions are fulfilled and then
stops.
Trigger mode REPEAT, AUTO
Recording and waveform evaluation is carried out continuously. Press the
STOP
key to terminate the measurement.
When "auto print" is set to ON, the waveform is printed out when operation
stops.
When "auto save" is set to ON, data are stored on media when operation
stops.
When memory segmentation (sequential save) is ON, data are stored in the
memory block only when operation stops.
Waveform evaluation consists of two actions, namely capturing data and
performing the evaluation. These two actions are carried out in sequence,
not simultaneously.
Therefore data are not captured while the evaluation is in progress, which
means that the input signal is not being continuously monitored. The time
required for evaluation is on the order of 750 ms.
If a high setting is chosen for recording length or if compression is used, the
evaluation cycle becomes slower.
On the waveform evaluation screen, A/B cursors can be used, but a partial
printout cannot be made.
When the waveform evaluation is specified, data equivalent to one screen
(30 divisions) is printed out. When waveform data having a recording
length of more than 30 divisions is to be printed out, the time axis should
be compressed. See Quick Start Section 8.3.