5.11 Setting the 8954 VOLTAGE/TEMP UNIT
────────────────────────────────────────────────────
Flashing cursor
Function
display
Meaning
:
The 8954 performs reference junction
compensation internally. (Measurement
accuracy is calculated from temperature
measurement accuracy and reference
junction compensation accuracy.)
:
The 8954 does not perform reference
junction compensation. Select this when
using an external reference junction
compensator. (Measurement accuracy is
tem
p
erature measurement accurac
y
onl
y
.
)
Flashing cursor
Function
display
Meaning
:
Discontinuity detection is not performed.
:
Discontinuity detection is performed.
(2) Setting reference junction compensation
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
RJC
item to be
set.
3. Use the function keys to make the selection.
(3) Setting the burn-out
When measuring temperature, check for
discontinuity in cord connecting the thermocouple
to the amp.
Normally, measurement values fluctuate when
measurements are taken with the cord
disconnected, but when this function is turned the
value displayed is the maximum of the values
obtained for measurement with discontinuities.
This waveform is attached to the uppermost part of
the display screen.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Burn Out
item to
be set.
3. Use the function keys to make a setting.